http://www.omega.come-mail: [email protected] SERIESPortable IR thermometers

Strany 2

102.5 KeyboardTraditional metal-click switches, with a working life of one million operations, are used to enter the operator’s instructions.The conta

Strany 3 - INTRODUCTORY NOTE

11In this case the instrument can be powered directly from mains using the external power supply module supplied with thethermometer.2.11 Logging Mode

Strany 4 - TABLE OF CONTENTS

123 PHYSICAL DESCRIPTIONOSP portable IR thermometers consist of a rugged, compact and self-extinguishing case ergonomically designed for aneasy prac

Strany 5

134 FUNCTIONAL DESCRIPTIONOSP series infrared portable thermometers block diagram is shown below: MeasuringobjectEmitted IR energyOptical systemFilter

Strany 6 - 1 GENERAL DESCRIPTION

144.3 MicrocontrollerThe microcontroller handles all the logic functions of the instrument, performs the linearization for non linear transducers,comp

Strany 7 - 1.1 Instrument codes

155 UNPACKINGRemove the instrument from its packing case and remove any shipping ties, clamps, or packing materials.Carefully follow any instruction g

Strany 8 - 1.2 Specifications

166 PRE-OPERATIONAL CHECKOSP series portable thermometers are powered either by four alkaline or by Ni-MH rechargeable batteries (optional).The extern

Strany 9 - 2 GENERAL FEATURES

177 POWER SUPPLYThe instrument may be powered by alkaline or rechargeable batteries or directly from the main line. OSP uses 4batteries type AA that a

Strany 10

188 WARNINGS & CAUTIONS8.1 Laser SightYou may receive harmful laser radiation exposure if you do not adhere to the warnings listed below:• USE OF

Strany 12 - 3 PHYSICAL DESCRIPTION

2OMEGAnetSM On-Line Servicehttp://www.omega.comInternet [email protected] North America:USA:ISO 9001 CertifiedOne Omega Drive, Box 4047Sta

Strany 13 - 4 FUNCTIONAL DESCRIPTION

209 OPERATIONSOSP series portable infrared thermometers are factory calibrated before shipment.During the start-up the operator should only select and

Strany 14 - 4.3 Microcontroller

21°Ce°Ce e°Ce°Ce°C e°Cfollowed, after few seconds, by one of the main operative display pages.The displayed page will be the same as when the instrume

Strany 15 - 5 UNPACKING

22In measuring mode, press the <*> key to enable (disable) the laser spot. After one second the message “LAS EN” (“LASDIS”) will be displayed.Wh

Strany 16 - 6 PRE-OPERATIONAL CHECK

23" In measuring mode, press the <Sel> key until the following page will appear:ee°CThe display indicates the actual temperature value tog

Strany 17 - 7 POWER SUPPLY

24e°C• Press the <SEL> key to obtain the following page that enables the High Alarm settinge°CHAL• Repeat the setting procedure used for the l

Strany 18 - 8 WARNINGS & CAUTIONS

25e°CLock e°CLock• When selected the required type of thermocouple press the <SEL> key to obtain the following page showingsimultaneously bo

Strany 19

269.9 Automatic Emissivity SettingThis function is used to make an automatic setting of emissivity using an external thermocouple connected with theau

Strany 20 - 9 OPERATIONS

2710 CONFIGURATIONThe operative configuration of the unit can be changed using the below indicated procedure.• Switch the instrument -Off-• Keep the

Strany 21 - 9.4 Laser sighting

28• Press the <SEL> key to select the new configuration step or press the <Trigger> key to switch the instrument -Off-.The new selection

Strany 22 - ALSO IN LOCK OPERATION MODE

2910.6 Buzzer On/OffIf the Alarms option is installed, it is possible to switch the buzzer on or off. This page allows the set-up of the acousticsigna

Strany 23 - 9.6 Alarm settings

3INTRODUCTORY NOTEThis user’s guide contains operating instructions, as well as a description of the principles of operation, of theOSP Series portab

Strany 24 - 9.7 Thermocouple input

30eSet°CeSet°C• Press the <▲> or <▼> key to enable (or to disable) the thermocouple auxiliary input operation.eSet°C• With the compensat

Strany 25

3110.11 Battery level of charge• Press the <SEL> key until the following page is displayed :eSet°CLock e+-Set°CLock eSet°

Strany 26

3211 DATA ACQUISITION OPERATIVE MODEWhen the optional accessory of the instrument is specified (Internal data memory + RS232 adapter + LogMan software

Strany 27 - 10 CONFIGURATION

33• Press the <MEM> key to start the acquisition.If automatic acquisition mode is set, OSP acquires automatically with the programmed time int

Strany 28 - 10.5 Acquisition settings

34eLog°CLock eLog°CLock eLog°CLock11.2 Recall stored data• Switch off the OSP thermometer.• Keeping pressed the <SEL> key, press the <

Strany 29 - 10.6 Buzzer On/Off

3512 LOGGING DATA MANAGERLogMan software allows OSP series users to set and prepare infrared thermometers to acquire the data organized by‘Tag’.This s

Strany 30 - 10.9 Firmware revision number

36These menu are subdivided in:FileNew, Open, Save, Print menu.NewCreates a new document.OpenOpens an existing document.OptionsSet Communication port,

Strany 31 - 10.11 Battery level of charge

3712.2.1 ToolbarsAdditional tool bars menu appear when the data view window or the graph view window is selected.Fig. 11.2a: Data view toolbarFig. 11.

Strany 32

38Copy into an open fileUse this command to copy selected data onto an opened file.It will appear a list of current opened files to be selected.This c

Strany 33

39" Select <OK> button to save the preference setting in the PC Memory." From the preference menu you can also set the header and foo

Strany 34 - 11.2 Recall stored data

4TABLE OF CONTENTS1 GENERAL DESCRIPTION...

Strany 35 - 12 LOGGING DATA MANAGER

40eLog°CLock• Press the <ENTER> key to clear the memory content and to obtain the following message that require confirmation :eLog°CLock• Pre

Strany 36

41Now you can save data on disk using name with extension LMD (for example TODAY.LMM).

Strany 37 - 12.2.1 Toolbars

4213 OPTIONS & ACCESSORIES13.1 Printer operations" Connect a compatible printer to OSP and switch the thermometer on.e°CLock" Press th

Strany 38 - 12.3 Quick starting

432 Remove the elevation and the windage adjustment caps.3 Switch on the two pinpointing lasers and aim them to a target placed at a distance equal

Strany 39

444 Screw up the polarized filter with the inserted optical filter placed on the front side of the sighting deviceCAUTIONIF YOUR OSP UNIT IS EQUIPPED

Strany 40

4514 APPLICATION NOTES14.1 Infrared energyEach body, at temperatures above the absolute zero (-273°C or 0K), emits energy in the form of electromagnet

Strany 41

4615 DIGITAL INTERFACEThe OSP portable IR thermometer can be equipped with a digital interface.The interface circuit is essentially based on the ser

Strany 42 - 13 OPTIONS & ACCESSORIES

47+ IBM 3 2 7 D B 2 5 IBM - PC+ 5 V 10 µF, 16 V 10 µF, 16 V 10 µF, 16 V + 158 Rx TxGroundIBM IBM TxRx6 2 16 10111214131 3 4 5 + + 10 µF, 16 V 10 µF,

Strany 43

48Reading values$00 - Pyrometer measure -------------------------------------------------------------- RXDATA1 STATUSDATA2 -DATA3 VALUE (H)DATA4 VALUE

Strany 44

49DATA4 'X' (ASCII CODE)$0C - Last 4 chars tag description -------------------------------------------------- RXDATA1 'X' (ASCII

Strany 45 - 14 APPLICATION NOTES

511 DATA ACQUISITION OPERATIVE MODE... 3211.1 How to operate in d

Strany 46 - 15 DIGITAL INTERFACE

50DATA4 VALUE (LL)SIGNED LONG VALUE = ALARM HI$85 - LAL Setpoint ------------------------------------------------------------------------ TXDATA1 VALU

Strany 47 - TSC 232

5128Ø PRINT "VALUE: " ; VALUE / 1Ø29Ø ENDExample B:1Ø CHAR = Ø2Ø IDNAME = 13Ø INSTRUCTION = &H804Ø VALUE = 805Ø VALUE$ = HEX$ (VALUE/100

Strany 48 - Reading values

5216 MAINTENANCEEach OSP portable infrared thermometer has been factory tested and calibrated before shipment.The calibration should be verified and

Strany 49

53APPENDIX

Strany 50 - 15.3.3 Communication programs

54A1 How to determine an object emissivityEmissivity is the measure of an object ability to absorb, transmit, and emit infrared energy. It can have a

Strany 51

55A1.2 Metals - Typical Emissivity Values1.0 µm 1.6 µm 5.1 µm 8-14 µmAluminumNon-Oxidized 0.1-0.2 0.02-0.2 0.02-0.2 0.02-0.1Oxidized 0.4 0.4 0.2-0.4

Strany 52 - 16 MAINTENANCE

561.0 µm 1.6 µm 5.1 µm 8-14 µmTungsten 0.1-0.6 0.05-0.5 0.03 Polished 0.35-0.4 0.1-0.3 0.05-0.25 0.03-0.1Zinc Oxidized 0.6 0.15 0.1 0.1

Strany 53 - APPENDIX

57INDEXAAcquisition by dedicated Tags; 32Acquisition settings; 28Alarm settings; 23Ambient Temperature Compensation (TAM); 29Analog Output; 10Analogue

Strany 54

58SSAFETY WARNINGS; 18Self calibration; 10Sighting telescope system; 42Specifications; 8Storage; 52TTABLE OF CONTENTS; 4Taking measurements; 9Target d

Strany 55

59WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13months from date

Strany 56

61 GENERAL DESCRIPTIONPortable infrared thermometers measure surface temperature without touching the surface. They collect the infraredenergy radiate

Strany 57

60Where Do I Find Everything I Need forProcess Measurement and Control?OMEGA…Of Course!TEMPERATURE!!!! Thermocouple, RTD & Thermistor Probes, Conn

Strany 58 - AFETY WARNINGS

71.1 Instrument codesTo OrderModel No. TemperatureRangeSpectralResponseTarget Spot Size@ Distance **D/S Optical RatioOSP500-(*)–30 °C to +930 °C-20°F

Strany 59 - WARRANTY/DISCLAIMER

81.2 SpecificationsTarget dimensions at different distancesNominal target diameter at 95% energy• Temperature measuring ranges:-30 to 2000°C (-22 to 3

Strany 60 - OMEGA…Of Course!

92 GENERAL FEATURES2.1 GeneralTemperature measurements of a liquid or gaseous compound have been successfully made with thermoelectric orexpansion the