http://www.omega.com e-mail: [email protected]®UserÕs GuideDP119-RTDMiniature Panel ThermometerMade in USA
3SECTION 2 ABOUT THE METER2.1 Front of the MeterFigure 2-1 shows the panel thermometer.Figure 2-1. Panel ThermometerFeatures:Display:3 1/2 Digit, 7-
42.2 Back of the MeterFigure 2-2 illustrates the rear of the meter.Figure 2-2. Rear View Showing P1, P2, P3 Locations2.3 Connector Description For
5SECTION 3 GETTING STARTED3.1 Main Board Power JumpersCaution: The meter has no power-on switch, so it willbe in operation as soon you apply power.
6Voltage Jumper Pin SettingsFigure 3-1 shows the location of solder jumpers W1 through W3.Figure 3-1. Power Jumper LocationsP3P1W1 W2 W321231TRANSFOR
73.2 Changing the Meter ParametersThe proper combinations of jumper plugs S1, S2, S3, S4, and S5inside the meter allow you to change various parameter
8The following tables summarizes what parameters you canchange.TABLE 3-1 Range Parameters* Refer to Table 1-1, for complete models (preface section).T
9Figure 3-3 gives an example of how to set S4 and S5 to configurea MF1 meter option. There is a jumper plug across A on S4 andanother jumper plug acro
10Section 3.3 Installation and Panel MountingFigure 3-4 shows the panel cutout dimensions, the dimensionsfor the panel thickness, and the backplate f
11Figure 3-4. Installation Information1/8 DIN PANEL CUTOUTBACKPLATESCREWSMETERBEZELLENSPANEL WITH1/8 DIN CUTOUT0.25 [6.4] MAX0.03 [0.8] MINPANEL THICK
123.4 Sensor Input ConnectionsThe RTD sensor (see Figure 3-5), can be wired in three differentways. They are 2-, 3-, and 4-wire configurations. The me
Servicing North America:USA: One Omega Drive, Box 4047ISO 9001 Certified Stamford, CT 06907-0047Tel: (203) 359-1660 FAX: (203) 359-7700e-mail: info@om
133.4 Sensor Input Connections (Continued)Figure 3-6. 2-Wire RTD WiringFigure 3-7. 3-Wire RTD WiringFigure 3-8. 4-Wire RTD WiringRefer to Table 3-1
143.5 Analog Output ConnectionsThe analog output is a linearized millivolt signal that is equivalentto the displayed temperature. It comes as a standa
153.6 Display Hold ConnectionsThe ÒDisplay Hold" function can be used to freeze the display.Connect a single pole, single throw (SPST) switch bet
163.7 Main Power ConnectionsConnect the ac main power connections as shown below.WARNING: Do not connect ac power to your meter untilyou have complete
17DC WIRINGThe meter can be configured at the factory to operate on a 9Ð26VDC non-isolated power. The DC voltage option is specified atthe time of ord
183.8 Disassembly/AssemblyYou must go through the following steps in order toremove the printed circuit board from the case if you wantto change from
19Figure 3-13. Removing the PC Board from the CaseWhen re-installing the printed circuit board:1. Hold the case (with the two mounting holes facing up
20SECTION 4 OPERATION AND CALIBRATIONImportant: These changes must be performed by aqualified technician.WARNING: Do not connect ac power to your me
214.2 CALIBRATION PROCEDURE1. Remove the front lens. Connect power to the meter,observing proper wiring procedures.2. Connect the precision resistance
22Figure 4-3. 4-Wire ConfigurationTABLE 4-1CALIBRATION VALUES WHEN YOU ARE USINGRESISTANCE SOURCE* Refer to Table 1-1, for complete models (preface se
TABLE OF CONTENTSSECTION PAGEPreface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .iiModels Available . . . . . . . . . . . . . . .
23SECTION 5 SPECIFICATIONSINPUT TYPE:RTD:...100-ohm platinum with alpha of 0.00385Calibration:...
24DISPLAY Type:...7-segment, LEDHeight: ...
TOP VIEWSIDE VIEW1.89[48.0]BEZELBACKPLATECASE3.78[96.0]3.76[95.5]0.76 [19.3]4.32[109.7]MAXPANEL CUTOUT0.25 [6.4] MAX0.03 [0.8] MINPANEL THICKNESS3.62
26SECTION 6 GLOSSARYCOMMON MODE REJECTION (CMR) - a measure of the affectof a voltage on the indicated display. It is the ratio of the commonmode vol
27NOTES
WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for aperiod of 13 months from date of
Where Do I Find Everything I Need for Process Measurement and Control? OMEGAÉOf Course!TEMPERATURE䡺⻬Thermocouple, RTD & Thermistor Probes, Connect
iiPREFACEManual Objectives: This manual shows you how to set up anduse the RTD Panel Thermometer.This meter is an economical Miniature Temperature Pan
iiiMODELS AVAILABLETABLE 1-1RTD MODELS AVAILABLEThe accuracies and ranges are listed in Section 5.TABLE 1-2Power Options AvailableMODEL POWERDP119-Mxx
ivTABLE 1-3Other Models AvailableThe following 3-1/2 digit mini thermocouple panelthermometers are discussed in a separate manual.MODEL TYPE ¡C or ¡F
vNOTES, WARNINGS and CAUTIONSInformation that is especially important to note is identified bythese labels:NOTE: provides you with information that is
1SECTION 1 INTRODUCTION1.1 UNPACKINGRemove the Packing List and verify that all equipment has beenreceived. If there are any questions about the s
21.2 SAFETY CONSIDERATIONSThis device is marked with the international caution symbol. It is important to read thismanual before installing or comm
Komentáře k této Příručce