Omega CL27 Uživatelský manuál Strana 44

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 46
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 43
42
47. Set both the U.U.T. and TC-simulator to output 32.0°F (Type T). Adjust
P1 for a null-meter reading of 0±3µV.
48. Hit OPR key to store T-calib zero adjustment.
NOTE: Steps 49 thru 54 apply to Model CL24 only. For Model
CL23A go next to step 55.
49. Select Type-E thermocouple.
50. Turn off U.U.T.
51. Replace T calibration-cable with E calibration-cable.
52. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
53. Set both the U.U.T. and thermocouple-simulator to 32.0°F (Type E).
Adjust P1 for a null-meter reading of 0±3µV.
54. Hit OPR Key to store E-calib zero adjustment
55. Turn off U.U.T.
56. Remove calibration cover. Be careful not to disturb battery connections.
57. Re-install J1 jumper.
58. Re-install original back cover. Calibration is complete.
Zobrazit stránku 43
1 2 ... 39 40 41 42 43 44 45 46

Komentáře k této Příručce

Žádné komentáře